SPLIT2 Real-time Dust Monitor

  • Accurate over a wide measuring range
  • Simple operation and instantaneous display of TWA, STEL, Minimum, Maximum in mg/m3
  • Can be used as an active monitor by combining with a sample pump
  • Operates for over 8 hours using rechargeable NiMH battery
  • Extended running time using AC charger
  • Compact and lightweight
  • Backlit display for use in dim light conditions
  • Audible alarm with adjustable threshold
  • Automatic data logging
  • Easy to use DustComm Pro Software for data downloading, trend analysis and comprehensive reporting of concentration changes over time
SKU: 770-300A Category:

Description

The Split2 is designed for personal or area monitoring for inhalable or respirable dust. It is a passive dust monitor that can become an active monitor/sampler by combining it with a 2 L/min sample pump and IOM sampling head or GS-3 Cyclone, for concurrent real-time dust monitoring and gravimetric dust sampling.

SPECIFICATIONS

Flow Rate 2 L/min using an external pump (not included)
Accuracy ±10% to NIOSH Method 0600 using SAE fine test dust
Sensing Range 0.01 to 200 mg/m3
Display Resolution 0.01 mg/m3
Particle Size Range 0.1 to 100 µm
Measurement Precision ±0.02 mg/m3
Calibration NIOSH gravimetric using NIST-traceable SAE fine test dust
Displayed Particle Concentration Instantaneous, TWA, STEL, Minimum, Maximum
Active Mode Use with sample pump at 2 L/min
Power Rechargeable NIMH battery
Operating Time ≥8 hours
Charging Time 12 hour
Digital Output RS-232
Data Storage 21,500 data points
Sampling Rate 1 second, 1 minute, 10 minutes
Temperature Operating: 0 to +50°C (32 to 122°F)
Storage:    -20 to +60°C (-4 to 140°F)
Dimensions Unit:     18 x 8 x 4.5 cm (7 x 3 x 2 inches)
Sensor: 4.4 x 3.8 cm (1.75 x 1.5 inches)
Weight 780 g with batteries

SPLIT2

Description: Split2 Monitor Kit, includes: Part Number
Monitor in carry pouch with detachable strap, detachable sensor with lapel clip, IOM sampling head, 110-240V AC Charger, calibration standard and zeroing filter, computer RS-232 cable, USB Memory Stick with instruction manual and DustComm Pro Software, factory calibration certificate 770-300A

ACCESSORIES

DOCUMENTS

Split-2-Real-Time-Dust-Monitor