OSHA Versatile Sampler (OVS) Tubes

The tubes combine sorbent and filter in one tube; a filter to trap aerosols and a two-section sorbent bed to adsorb vapours in one specially constructed glass tube that eliminates filter and tube sampling trains. Aerosols and vapours can be collected simultaneously. Ideal for sampling pesticides, explosives and glycols. Only cleaned and verified materials are used in OVS tubes to ensure low background interference. A flow rate of 1 L/min is typically used, and the samples are solvent extracted and analysed by gas chromatography (GC) or high performance liquid chromatography (HPLC) with detector.

  • Sorbent and filter combined in one tube for NIOSH and OSHA
  • Collect vapors and aerosols simultaneously
  • Available with a variety of sorbents for special uses
  • Low backgrounds ensure sample integrity
  • Meet OSHA and NIOSH design specifications

For mixed-phase sampling using a true inhalable sampler that meets ISO 7708/CEN, see the IFV Pro Sampler.

The size of all the OVS tubes is 13→8 x 75 mm (Length x OD) and the tubes have open glass ends. The OVS Tube Holder (224-29V) is designed to accommodate the 13 mm diameter of SKC OVS tubes, provide a convenient clip to attach the tube in the breathing zone, and protect the tube during sampling.

OVS Tubes

OVS Tubes

          OVS Tubes

OVS Tubes Construction

Note: All OVS tubes have limited shelf-life; refrigerator/freezer storage may be required. Contact SKC for more information.

To select a tube for a specific compound refer to the SKC Sampling Guides or use the SKC online Hazard Database

Description

Application Method Sorbent (mg) Separators Pack Size Part Number
Pesticides

Organotin Compounds *  

OSHA 62, 63, 67, 70, 74, OSHA CSI

OSHA CSI

XAD-2 (270/140)  FFGT 10

50

226-30-16

226-30-16A

Pesticides, organophosphorus NIOSH 5600, 5601, 5602 XAD-2 (270/140)  FFQT 10

50

226-58

226-58A

Explosives
(trinitrotoluene (TNT) and dinitrotoluene (DNT))

Phthalate esters

Acrylates and Benzophenone

OSHA 44

OSHA 104

Non-agency

Tenax TA (140/70)  FFGT 10 226-56
Caprolactum Vapour

Glycols

OSHA CSI

NIOSH 5523

XAD-7 (200/100)  FFGT 10

50

226-57

226-57A

Kathon® 886 Biocide Non-agency Silica Gel (520/260)  FFGT 10 226-99

Table Notes:

* Methyl tin mercaptide, stannous-2-ethyl hexanoate, butyltin trichloride

Separators:

G:  Glass Fibre Filter
F:   Foam
Q:   Quartz Filter
T:   PTFE Ring

Accessories

OVS Tube Holder

Includes fitting with special durable protective cover, 0.9 m (3 ft) tubing and collar clip – Part Number 224-29V

 

OVS Tubing/Adapter Kit

Includes 2 sections of tubing and tubing adapter. Used to facilitate calibration of a sample pump when using an OVS tube. The tygon tubing is 1/4-inch ID x 3/8-inch OD, the silicone tubing is 3/8-inch ID x 5/8-inch OD. – Part Number 224-31